|
|
||||||||
Research Reports |
F Manning, MD, PT, is Family Medicine Resident, Faculty of Medicine, University of British Columbia, Vancouver, British Columbia, Canada. This study was completed in partial fulfillment of the requirements for Dr Manning's Master of Science degree
E Dean, PhD, PT, is Professor, School of Rehabilitation Sciences, University of British Columbia, T325-2211 Wesbrook Mall, Vancouver, British Columbia, Canada V6T 1Z3 (elizdean{at}rehab.ubc.ca). Address all correspondence to Dr Dean
J Ross, PT, is Section Head, Critical Care, Rehabilitation Services, Vancouver Hospital, Vancouver, British Columbia, Canada
RT Abboud, MD, FRCPC, is Professor, Division of Respiratory Medicine, Faculty of Medicine, University of British Columbia, and Director, Lung Function Laboratory, Vancouver Hospital and Health Sciences Centre, Vancouver, British Columbia, Canada
Submitted December 4, 1997;
Accepted January 11, 1999
| Abstract |
|---|
Key Words: Body position Diffusing capacity Older adults Oxygen transport Pulmonary function Side lying Recumbency Ventilatory inhomogeneity
| Introduction |
|---|
|
|
|---|
Although side-lying positions are commonly used clinically, the differential effects of right and left side lying on lung function compared with a reference position such as upright sitting have not been studied in detail. There have been a few reports of improved arterial oxygenation in left versus right side lying in patients with unilateral lung disease35 and bilateral lung disease6 and in patients following coronary artery bypass surgery.6 In recumbent positions, gas exchange is improved with the healthy lung down in patients with unilateral lung disease3 and in right side lying in patients with bilateral lung disease.5 In patients with unilateral lung disease, the role of the inferior lung as a gas exchanger is enhanced because of the cephalad displacement of the hemidiaphragm placing it at a greater mechanical advantage.1 In addition, the expansive forces on the superior lung that maximize gas exchange in that lung may also contribute. In patients with bilateral lung disease, gas exchange may be enhanced due to the increased volume of the right lung anatomically and less effect of cardiac compression on this lung.1
The purpose of our study was to replicate and extend the existing body of knowledge pertaining to the normal relationship between side lying and several lung function variables in older age groups. Given that the majority of hospitalized patients tend to be older and our belief that physical therapists need to understand normal responses as a basis for understanding abnormal responses superimposed by pathology, we studied individuals aged 50 years or older with no known history of cardiac or pulmonary disease. The bony structure of the thorax, the ventilatory muscles, the lung parenchyma, the pulmonary vasculature, and the heart are affected by the aging process.7,8 The interrelated changes in these structures result in a gradual decline in cardiopulmonary function and gas exchange with aging that is distinguishable from the more profound loss of function that occurs as a result of disease.79
The results of our study, therefore, could yield greater understanding of the relationship between the side-lying positions and lung function, which is fundamental to (1) improving our understanding of this relationship when pathophysiology is superimposed, (2) using body positioning as a primary intervention to maintain or improve overall gas exchange in patients with cardiopulmonary dysfunction, and (3) minimizing the effects of deleterious body positions. Thus, the findings of this study will refine the principles for prescribing body positioning as opposed to routine body positioning.
| Method |
|---|
|
|
|---|
Analysis of the slope of phase III, the expired alveolar plateau, of the SBN2 test has been used for many years to measure the uniformity or homogeneity of ventilation in the lung, a requisite for optimal oxygenation.10 The subject's exhalation is analyzed for nitrogen (80% of the composition of normal air that we breathe) after a full inspiration of 100% oxygen from residual volume. The first part of the exhaled gas from the anatomic dead space has a 0% concentration of nitrogen. Then, there is a sharp rise in nitrogen concentration as gas from the alveoli becomes mixed with gas from the dead space. This is followed by a relative plateau as more gas is emptied from the alveoli. A steep slope of this phase indicates nonuniform distribution of alveolar gas. The greater the nonuniformity or inhomogeneity of ventilation, the greater the ventilation to perfusion mismatch, hence, deoxygenation.
In each position, spirometric tests were conducted first to provide a reference value for vital capacity (VC) for the test of DLCO. The test of DLCO was conducted before the SBN2 test because the latter test involved inhaling 100% oxygen, which could influence the results of the DLCO test. Each session lasted approximately 2
hours. Subjects rested between tests and between repeated trials of each test.
Subjects
Nineteen subjects (11 women, 8 men), with a mean age of 62.8 years (SD=6.8, range=5074), participated in the study. Nine subjects were lifetime nonsmokers, and 10 subjects were ex-smokers. The average height of the women was 159 cm (SD=6, range=145168), and that of the men was 176 cm (SD=8, range=163183). The average weight of the women was 56.6 kg (SD=8.2, range=4576), and that of the men was 72.0 kg (SD=11.0, range=5693). All subjects had no known history of cardiac or pulmonary disease. The group was recruited from the university community through a public announcement.
Procedure
We studied the following dependent variables in different body positions: (1) pulmonary function, including spirometric measures of FVC and FEV1, (2) inhomogeneity of ventilation, as measured by the slope of phase III (DN2%/L) of the SBN2 test, and (3) pulmonary diffusing capacity, based on the single-breath test adjusted for alveolar ventilation (DLCO/VA). The spirometric variables were included for the purposes of screening for airway obstruction, setting a baseline for acceptability of the DLCO test (which requires the subject to inhale a minimum of 90% of his or her VC), and evaluating changes between different positions. The DN2%/L and DLCO/VA provide valuable information with respect to the capacity of the lung to serve as a gas exchanger. The DN2%/L focuses on the process of ventilation and its distribution (ie, the movement of inspired gas from the atmosphere to the alveoli). The DLCO/VA focuses on the process of diffusion and involves the transport of gases across the alveolar-capillary membrane. The distribution of ventilation reflected by the homogeneity of ventilation within the lung, in combination with the distribution of pulmonary perfusion, is a determinant of the efficiency of gas exchange. All spirometric and lung function indexes were obtained according to the American Thoracic Society (ATS) standards.1113 All subjects were asked to refrain from vigorous exercise prior to testing on the day of the test session. They were also requested to avoid eating a heavy meal within 2 hours of the test and to wear comfortable, nonrestrictive clothing.
The general experimental procedure is outlined in the Figure. The protocol for session A was randomly selected for the first subject on the first visit, and the protocol for session B was used on the second visit. For successive subjects, the 2 protocols were alternated. On arrival at the laboratory for the first session, the test procedures were explained to the subjects, who then gave written consent to participate in the study. The time taken for this discussion and for the determination of individual heights and weights allowed for a rest period prior to testing. It also allowed for familiarization with the environment and the tester. In both sessions, the first test position was sitting and the second test position was either left or right side lying. The order of lung function tests, as shown in the Figure, was the same for each visit, as necessitated by spirometric prerequisites for DLCO and the potential for the 100% oxygen inhaled during the SBN2 test to interfere with the results of the DLCO test. Both testing sessions were completed within 1 week.
|
Subjects assumed the test position 15 minutes prior to the first trial of either the DLCO test or the SBN2 test in each position. This rest period was included to accommodate the effects of position change on the pulmonary circulation, notably the pulmonary capillary blood volume, which are known to be time-dependent.14,15 Although the time course and duration of position-related changes in the pulmonary circulation are not well-defined in the literature, various studies of the effect of position change on DLCO have used rest periods of 15 minutes or less.1618
Measurements from the SensorMedics 2200 System.
The SensorMedics 2200 System* has both a nitrogen analyzer, which is used in the SBN2 test, and a rapid response multi-gas analyzer, which is used in the test of DLCO.19 For measuring the nitrogen concentration of expired gas during the SBN2 test, a needle valve is mounted into a mouthpiece adapter. The valve is connected to a fast-response nitrogen analyzer containing an ionization chamber. Maximum nitrogen ionization is achieved by an optimum negative pressure (created by a vacuum pump) in the analyzer. When nitrogen ionizes, it emits ultraviolet light, the intensity of which is directly related to the concentration of nitrogen. The light energy is converted into an electric signal that is translated by the computer into a concentration. The response time for this analysis is given as less than 50 milliseconds. A different method is used for measuring the carbon monoxide concentration of expired gas during the single-breath DLCO test. In this case, the measuring principle is the nondispersive infrared absorption technique. The expired gas is exposed to a beam of infrared energy and absorbs a certain amount of this energy, depending on the partial pressure of the gas (which, in turn, is dependent on the concentration of the gas). The infrared absorption is measured and converted to an electric signal that is relayed to the computer.
Calibration for flow and volume measurements was done prior to each session and according to the procedure outlined in the SensorMedics 2200 operator's manual.19 This procedure involved using a 3-L calibration syringe to deliver room air into the system. Calibration of the nitrogen analyzer was performed a least once a week, in accordance with the terms outlined in the operator's manual. This calibration involved "peaking the needle" by opening or closing the needle valve (to allow more or less gas flow into the analyzer) to obtain the maximum percentage of nitrogen reading on room air. The vacuum pump was turned on a minimum of 20 minutes prior to calibration, in accordance with the specified calibration procedure. A linearity check of the nitrogen analyzer was performed at the start of the study and twice during the study. This check involved exposing the SensorMedics 2200 to 5 different known concentrations of nitrogen (ranging from 0.00% to room air at 79%; 100% oxygen [0% nitrogen] to air [79% nitrogen]) for analysis.
Depending on the specific test selected from the computer menu, a valve automatically switched the subject, who was connected to the system by a breathing tube and mouthpiece, between room air, 100% oxygen, or a multi-gas mixture. Throughout a test, tracings on the computer screen provided visual feedback for the subject and coaching cues for the tester. In addition, "best" results for spirometric tests (as per ATS standards) and averages for DN2%/L and DLCO were displayed, as were the values expressed as a percentage of the predicted values. For our study, we used the prediction equations of Crapo et al20 for volume measurements, those of Knudson et al21 for flow measurements, those of Buist and Ross22 for DN2%/L measurements, and those of Miller et al23 for diffusion measurements.
Performance of spirometric tests.
The maximum expiratory maneuvers were conducted according to ATS standards.11 Positioned in either sitting or side lying, the subject was monitored throughout the test to prevent alterations in body position. Following a detailed explanation of the test, the subject was connected to the mouthpiece with the noseclip in place. After several tidal breaths, the subject was coached through the maximal forced expiration procedure. The end of the test, as defined by ATS criteria of no change in volume for at least 2 seconds following an exhalation time of at least 6 seconds, was indicated by a computer message displayed on the screen. The subject expired with maximal expiratory effort until 3 acceptable tracings were recorded (which involved a maximum of 6 trials). In accordance with ATS recommendations for test reproducibility, the 2 largest FVCs (taken from acceptable curves) varied by less than 5%. Similar reproducibility criteria were applied to the measurement of FEV1.
From the repeated trials that met ATS criteria,13 the largest FVC measurement and the largest FEV1 measurement were selected for data analysis. The selected spirometric values were compared by the computer with reference values and were presented as a percentage of the predicted value.
Performance of the SBN2 test.
The SBN2 tests were conducted according to the National Heart and Lung Institute standards.12 Following a detailed explanation of the maneuver, the subject was connected to the mouthpiece with a noseclip in place. After several tidal breaths, the subject inhaled 2 slow, deep breaths and then exhaled completely. From residual volume (which is recognized by the computer as the point at which there is no expiratory flow for 0.5 second), the subject inhaled a slow VC breath of oxygen. The flow rate of the subsequent slow exhalation was maintained between 0.3 and 0.6 L/s for as long as possible. To assist the subject in maintaining a flow rate in this range, a resistance was inserted in the expired circuit to slow down expiratory flow and the flow rate was displayed on the computer screen during exhalation, along with the upper and lower limits of the flow rate range.
In each test position, the SBN2 test was repeated until 3 acceptable tracings were obtained according to the National Heart and Lung Institute standards.12 Rest periods of approximately 5 minutes were given between trials to provide time for washout of excess oxygen and thus the restoration of the normal nitrogen gradient in expired air. The DN2%/L was calculated by the computer from analysis of the final expirate, using the increase in nitrogen concentration over 1 L of the expired volume, between 750 mL (representing the complete dead space washout) and 1,750 mL. The average DN2%/L value from 3 acceptable tests was used for data analysis and was expressed as a percentage of the predicted value.
Performance of the DLCO test.
The DLCO tests were conducted according to ATS standards11 for a single-breath test. As with the other tests, the subject was connected to the mouthpiece with a noseclip in place for a short period of tidal breathing before the test maneuver. On a cue from the tester, the subject exhaled to residual volume and then quickly inhaled a VC measure of the test gas, which contained minimal, nontoxic concentrations of carbon monoxide (0.30%), methane (0.30%), and acetylene (0.30%); 20.95% oxygen; and the balance, nitrogen. On the basis of previous forced expiratory maneuvers, which gave a measure of VC in the same position, the computer displayed the minimum acceptable volume (at least 90% of the subject's VC) by a horizontal target line, which the subject could see while inhaling. The subject was coached to hold his or her breath at total lung capacity for 10 seconds, during which time he or she was encouraged to relax against the closed valve of the unit. The end of the breath hold was signaled by the crossing of the volume versus time tracing over a vertical time line, after which the expiratory valve automatically opened and the subject was coached to exhale rapidly and completely. The precise breath-hold time was calculated by the computer and based on ATS standards. The mean concentrations of methane and carbon monoxide were calculated by the computer over a 1,000-mL alveolar sample volume from 750 to 1,750 mL of the expired volume. The sample collection volume was not adjusted larger than 1,000 mL.
In each test position, the DLCO test was repeated until at least 2 acceptable tracings were obtained. A rest interval of at least 4 minutes was given between trials to allow the test gas to wash out from the subject's lungs. Before each trial, an automated procedure was initiated in which methane and carbon monoxide concentration readings were zeroed with room air. The 2 tests from which an average DLCO/VA was calculated were within 10% of each other, in accordance with ATS recommendations.11 The average DLCO/VA was expressed as a percentage of the predicted value for data analysis. Although an adjustment for hemoglobin is desirable in the calculation of DLCO, it is not considered mandatory by the ATS11 for subjects with no known history of cardiac or pulmonary disease and was not made in this study. In addition, based on the findings of Pistelli et al,18 we assumed that the hemoglobin concentration was constant for every subject throughout the study. Because the 2 test sessions were conducted within 1 week, the effect of normal fluctuations in hemoglobin on DLCO was presumably minimized. Thus, we were confident that hemoglobin changes did not confound DLCO when different positions were studied.
Data Analysis
Descriptive statistics were calculated for age, height, and weight. Descriptive statistics were also calculated for the 4 dependent variables (ie, FVC, FEV1, DN2%/L, and DLCO/VA). Individual subject test results were compared with predicted values. For each subject, the observed value was compared with the lowest acceptable normal limit (for the spirometric variables and for DLCO) or the highest acceptable normal limit (for DN2%/L). The limits of normal were obtained by subtracting (for the lowest limit) or adding (for the highest limit) the 95th percent confidence interval (1.65 x standard error of the estimate) from the predicted value for each subject. Coefficients of variation (CVs) were calculated to establish the reliability of the measurements.
Within-subject one-way analyses of variance (ANOVAs) were used to analyze the effect of body position on each of the dependent measures. Post hoc comparisons were made using the Duncan multiple range test. A significance level of P<.05 was selected for all statistical tests.
| Results |
|---|
|
|
|---|
|
|
|
| Discussion |
|---|
|
|
|---|
The decrease in FEV1 in a side-lying position compared with a sitting position is in agreement with the relatively few studies demonstrating changes in FEV1 with recumbency.17,18 Although direct comparison with these studies cannot be made because the effects of the recumbent position appear to vary between side-lying and supine positions, based on the results of our study, the similarity of results suggests that recumbent positions also limit expiratory volumes and flow. The exact cause of this apparent obstructive process, which may reflect an increase in airway resistance, a decrease in elastic recoil of the lung, or decreased mechanical advantage of forced expiration, presumably affecting the large upper airways, is not clear. In our study, care was taken to position subjects so that the head and neck were in alignment with the trunk and in a neutral position (in both sitting and side-lying positions). This neutral position was adopted to avoid extraneous stress on the upper airways. Anthonisen32 reported that hyperextension of the neck results in an increase in FEV1, secondary to elongation and stiffening of the trachea, thereby facilitating air flow. Conversely, the relaxed recumbent positions examined in our study may effectively shorten and increase the compliance of the airways, such that these positions limit rather than augment maximal forced expiratory effort.
The slope of phase III, DN2%/L, of the SBN2 washout test reflects both interregional and intraregional differences in the lung.2 Even in health, the CV for DN2%/L is well-known to be high compared with other indices of lung function.24 The mechanism underlying the greater ventilatory inhomogeneity and the higher CV for DN2%/L in a side-lying position compared with a sitting position could reflect several factors that affect the distribution of ventilation, especially in older people. Key contributory factors include airway closure and increased pulmonary time constants, both of which are adversely affected in recumbent positions as well as with advancing age.33 Anatomical factors such as an increase in the weight and volume of the heart with aging,34 and thus greater impingement on adjacent lung parenchyma in the left side-lying position, may also have had a role. Whether residual effects from a history of smoking affected intraregional differences in 10 subjects is unclear.
Statistically, our results do not support increased inhomogeneity of ventilation in a side-lying position, which would be predicted for an older age group. As described by Otis et al,35 the concept of pulmonary time constants explains the variations in gas entry into independently ventilated lung units. Increased pulmonary time constants, caused by regional changes in airway resistance and compliance and leading to varying degrees of filling of lung units, can increase the inhomogeneity of ventilation.36 From determinations of lung resistance and lung compliance in different body positions, Behrakis et al30 concluded that pulmonary time constants overall were greater in a side-lying position than in a sitting position in young adults without cardiopulmonary impairments. This conclusion was based on the disproportionate increase in lung resistance in a side-lying position (40% greater than in a sitting position) compared with the decrease in lung compliance (10% less than in a sitting position). Thus, the product of resistance and compliance (ie, the pulmonary time constant) was greater in a side-lying position than in a sitting position. Michels et al37 compared supine and sitting positions and found a similar position-dependent increase in resistance of the respiratory system in a group of adults without cardiopulmonary impairments aged 20 to 67 years. Furthermore, they found that, in a subgroup of young subjects, the increase was more marked in smokers than in nonsmokers. In addition, the increase was more marked with aging, such that over the age of 50 years, both nonsmokers and smokers demonstrated position-dependent increases in resistance that were of the same magnitude. Behrakis et al30 suggested that changes in geometry of the upper airways, the aperture of the glottis, or both may contribute to this effect. Michels et al37 proposed that intrinsic narrowing of the peripheral airways of smokers may be more pronounced in recumbency than in a sitting position.
In light of these considerations, we would anticipate that, in comparison with the sitting position, both left and right side lying would reflect the effects of airway closure and increased pulmonary time constants in conjunction with increased inhomogeneity of ventilation. The fact that our results do not support the findings of other researchers suggests that the effect of body position on the inhomogeneity of ventilation is more variable than for the spirometric measures. Rather than reflect on the results of "outliers," our results indicate that age-related factors could have had a role. The large variability of the DN2%/L may have obscured a real difference between the left and right side-lying positions. Further study is needed to elucidate differential effects between left and right side-lying positions in older people and whether this effect is accentuated further with factors such as pathology, smoking history, and obesity.
The effect of recumbency on diffusing capacity is conflicting in the literature and may reflect age-related factors. The lack of a difference in DLCO/VA between sitting and either left or right side lying in older subjects in our study is consistent with the work of Stam et al,38 although they used supine rather than side lying as the recumbent position. These findings are in contrast to those of investigators3941 who found increases in DLCO/VA of up to 15% from a sitting position to a recumbent, usually supine, position. This discrepancy may be explained by a greater unevenness of DLCO/VA through the lung in a side-lying position than in a supine position (due to an increased transverse diameter of the chest compared with anteroposterior diameter and the greater vertical gradient in a side-lying position than in a supine position). Furthermore, few investigators have targeted an older population for the study of position-related changes.
Some authors have reported that DLCO decreases with increasing age. Georges et al9 attributed the decrease to a decline in Dm (the membrane component of DLCO) after the age of 40 years and to a decrease in pulmonary capillary blood volume after the age of 60 years. In addition, the anatomic changes associated with aging that affect Dm and pulmonary capillary blood volume may account for the apparent differences in response to a position change as compared with a young population. Brody and Thurlbeck,33 for example, described a loss of alveolar surface area, a possible decrease in number of pulmonary capillaries, and an increase in the inner diameter of alveoli, which may affect the mixing of gases by diffusion, as morphologic changes accompanying aging. These changes may reduce the increase of pulmonary capillary blood volume from a sitting position to a recumbent position and may be reflected by a constancy of DLCO between positions. Further study of underlying mechanisms, however, is needed in light of reports that there is no difference in the distribution of pulmonary perfusion between older and younger individuals and little change in pulmonary capillary density.42 Similarly, further study is needed to explain these discrepancies in the literature.
Effect of Left Versus Right Side-Lying Positions
The similarity of the results for left side lying and right side lying are in agreement with the results of the study by Behrakis et al,30 one of the few studies comparing lung function in the side-lying positions. These investigators also reported no differences between the side-lying positions for VC, expiratory reserve volume, static and dynamic compliance of the lung, resistance of the lung, or pulmonary time constants in a young group of subjects. In other studies of the side-lying positions, however, there was either the selection of only one of the side-lying positions, usually right side-lying,4346 or no reported differentiation between left and right side-lying positions.4749
In younger adults without cardiopulmonary impairments, there is no reason to suspect a difference in lung function between left and right side-lying positions. In an older population, however, the age-related variation in cardiopulmonary status (eg, increase in weight and volume of the heart,7,8 changes in mediastinal compliance) may result in differences in function between left and right side-lying positions when compared with a sitting position. Furthermore, in the presence of cardiac or pulmonary conditions, position-related effects on cardiopulmonary status may be accentuated in left and right side-lying positions. Zack et al,5 for example, reported that arterial oxygen tension was generally higher in right side lying than in left side lying in 13 patients (ages not reported) with equally distributed bilateral lung disease, whereas there was no difference in arterial oxygen tension between side-lying positions in 6 control subjects (mean age=25 years). This effect was attributed to the smaller volume of the left lung and compression of the heart on the left lung in left side lying. The morphological changes in the lung that are associated with aging could have a similar effect if the changes are unequally distributed between the left and right lungs; however, there is no evidence to suggest that this is the case. Regional age-related changes in the lung have not been reported, except for a greater degree of emphysematous change (considered as a normal part of the aging process in nonsmokers) in the lower zones compared with the upper zones of the lung.33
Clinical Implications
Physical therapists should anticipate the physiological effects of side-lying body positions when managing their patients, who may assume such positions for comfort and rest, may be placed in these positions to avoid the negative effects of static body positions, or may be placed in specific therapeutic body positions to augment arterial oxygenation or drain pulmonary secretions. Based on our results, we contend that physical therapists should consider the predictable reduction in lung volumes, lung capacities, and flow rates when placing patients in side-lying positions. Furthermore, based on a comparison of findings in the literature, we believe that therapists should also consider the less predictable changes in diffusing capacity and homogeneity of ventilation.
The adverse effects of recumbent positions can be accentuated by the following factors: cardiopulmonary pathology, age, smoking history, obesity, breathing at low lung volumes, sedation, anesthesia, oxygen, and other pharmacological agents.2 Thus, patients who may be scheduled for even minor medical or surgical procedures may be at risk because of these confounding factors. We believe it is essential that physical therapists be able to identify patients who are at risk to ensure that they are appropriately monitored and that upright positions are encouraged over recumbent positions as much as possible. In our view, therapeutic recumbent positions should be prescribed judiciously between treatments, and early intervention should be instituted if necessary. In addition, body positioning should not be used injudiciously or without appropriate monitoring, even when administered routinely for comfort and for avoidance of the negative effects of prolonged static positioning. Without due precautions, a patient who is apparently at low risk can readily become a patient who is at high risk.
Finally, of considerable clinical importance is the fact that a greater physiologic understanding of the strong and direct effects of body positioning on cardiopulmonary function and gas exchange has helped to explain both improvements previously attributed to conventional "chest" physical therapy and the increasing number of negative outcomes in response to this time-honored treatment. We believe that the results of studies evaluating interventions for patients with cardiopulmonary disorders (eg, measures of lung function and gas exchange) have been confounded by the effects of body positioning (or body positioning in combination with mobilization), irrespective of the effect of airway clearance.6 Because optimal lung function is associated with the upright positions,2 the results of our study lend further support to exploiting the therapeutic benefits of these positions.
| Conclusions |
|---|
|
|
|---|
The absence of any change in diffusing capacity in older people without cardiopulmonary impairments supports the idea that side lying may not induce the comparable changes in pulmonary capillary blood volume and venous return that are reported in the literature and that are responsible for the associated increase in diffusing capacity in a supine position. Although no increase in inhomogeneity of ventilation was observed in our study, this finding does not minimize the well-known detrimental effects of recumbency on functional residual capacity and associated arterial desaturation.
Our results have implications for both routine and therapeutic positioning of hospitalized patients with or without cardiac or pulmonary conditions. The physical therapist needs a thorough knowledge of all factors that affect all steps in the oxygen transport pathway in order to prescribe this body positioning efficaciously (ie, with maximal benefit and least risk).
| Acknowledgments |
|---|
| Footnotes |
|---|
This study was supported, in part, by funding from the Canadian Lung Association.
* Summit Technologies Inc, 84071H Ave SW, Ste 900, Calgary, Alberta, Canada T2P 3G2. ![]()
| References |
|---|
|
|
|---|
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |