|
|
||||||||
Technical Reports |
P Segers, PhD, is Post-doctoral Researcher, Hydraulics Laboratory, Institute Biomedical Technology, Ghent University, Sint-Pietersnieuwstraat 41, 9000 Ghent, Belgium (patrick.segers{at}navier.rug.ac.be). Address all correspondence to Dr Segers
JP Belgrado, PT, is Researcher, Service de Kinésitherapie et de Readaptation, Université Libre de Bruxelles, Avenue FD Roosevelt 50, B-1050 Bruxelles, Belgium
A Leduc, PhD, is Professor, Service de Kinésitherapie et de Readaptation, Université Libre de Bruxelles
O Leduc, PhD, is Post-doctoral Researcher, Service de Kinésitherapie et de Readaptation, Université Libre de Bruxelles
P Verdonck, PhD, is Professor, Hydraulics Laboratory, Institute Biomedical Technology, Ghent University
Submitted February 27, 2002;
Accepted April 28, 2002
| Abstract |
|---|
Key Words: Breast cancer Compression therapy Deep venous thrombosis Lymphedema Mastectomy Therapy
| Introduction |
|---|
|
|
|---|
Not all authors agree on whether compression therapy, called "pressotherapy" in some parts of the world, should be part of the treatment.7 We contend that this lack of agreement is probably due to the lack of references concerning optimal pressure values and how to apply compression. Methods such as the application of pressure on a limb that are necessary for efficient emptying of the venous or lymphatic system without increasing the heart load or damaging the vascular parenchym8 are not defined and are in need of scientific foundation and standardization. The optimal pressure for lymphatic drainage to which each chamber of a sequential pressure device is inflated is not clear and may depend on the application of the pressure cuff. Generally, pressures in a relatively wide range between 30 and 100 mm Hg are used.911 During pressure application, patients are usually in a supine position and the pressure in the superficial veins can be assumed not to exceed 20 mm Hg. As such, occlusion of the superficial venous network should be obtained at a compression of about 40 mm Hg, and it can be expected that the application of higher pressures will not further empty the superficial venous network. Pressure over 40 mm Hg raises the load on the heart through an elevation of the intra-auricular and pulmonary capillary pressure.12 This is not the purpose of compression therapy.
Periodic external compression for emptying the venous system of the limb is best performed in a nonuniform, graded way, with maximal pressure at the extremity of the limb.9,1316 This can be achieved by using specially designed cuffs, consisting of multiple chambers, which are inflated in a sequential order from the extremity of the limb toward the torso. Compression devices, in our experience, have simple controls, and only the air pressure inside their internal air reservoir is controlled. Cuff chamber pressures, in our experience, are not monitored during the entire inflation sequence of the cuff and the pressure exerted on the patient's limb is not measured. We conducted this study for 2 purposes. First, we wanted to verify whether the pressures applied in the cuff chambers are actually transferred to the patient's limb. Second, we wanted to monitor the pressures in different cuff chambers during inflation and to assess the interaction of adjacent cuff chambers and the consequences for the pressure control of these pneumatic devices.
| Materials and Methods |
|---|
|
|
|---|
|
A T-junction was inserted in the tubes connecting the pressure controller to the cuff chambers (Fig. 1). This allowed us to use piezoelectric pressure transducers (DTX/Plus Pressure Transducer
) of the type routinely used in hospitals for cardiovascular pressure measurements. We continuously monitored the pressure inside the cuff chamber (Pchamber). To measure the contact pressure between the model and the compression cuff, we filled 500-mL infusion bags with about 50 mL of water and deaerated the bags by manually squeezing out the air. A connector was glued into the bag to allow coupling to a 3-way stopcock and the piezoelectric pressure transducers. When inflated, the cuff pressure is transferred onto the infusion bag, and the pressure measured inside the bag is thus representative of the pressure exerted by the cuff on the cuff-model interface (Pinterface). Four of these bags were attached to the limb models at different measuring locations (Fig. 1), and the instrumented model was then inserted into the cuff.
The maximum deviation of the pressure transducers, due to the total effects of nonlinearity, hysteresis, and sensitivity variation, is no more than 2% of the reading (minimum value of 1 mm Hg) according to the product specifications provided by the manufacturer. The pressure transducers were calibrated before and after each measurement series by connecting the sensors to a water column, with water level controlled over a range of 0 to 115 cm (corresponding to pressures of 085 mm Hg). No drift of pressure transducer offset or gain was observed. The response of the transducers is high (natural frequency in the order of 200 Hz), but the frequency response of the complete measurement setup strongly depends on the properties of the tubes (length, stiffness, diameter, and density of medium in the tubes) connecting the transducer to the location where pressure is being monitored. In our study, the transducers were connected directly to the infusion bag (to measure Pinterface) or to the T-junction (Pchamber) on the pressure line. When measuring Pchamber, air is present in the connection to the pressure transducer, lowering the frequency response of the system. Our main interest, however, was the (quasi-static) maximum pressure level reached within the cuff or at the cuff-model interface, which depends on the low-frequency contents of the signal only.
We were able to simultaneously capture the analog signals of 4 pressure transducers (National Instruments A/D card
). Data were sampled at 5 Hz and visualized on a computer screen using customized software (Labview).
Measurement Protocol
Data were gathered for preset working pressures of 30, 60, 80, and 100 mm Hg. Preset working pressures were the same for all chambers. For the 3 cylindrical models, we measured both Pchamber and Pinterface in a successive way. For a given working pressure, Pchamber was first measured in the 4 most distal chambers. The pressure transducers were then connected to the infusion bags, and the cuff was inflated. The model was placed horizontally on a laboratory table.
For measurements with the ellipsoidal model, the model was placed in a vertical position. Four infusion bags were attached to the 4 vertices of the ellipsoidal model at the same height in order to eliminate gravity pressure differences among the 4 sensors. Only Pinterface was measured. The model was inserted into the cuff, with the measurement location under chamber 2. The model was moved such that the instrumented bags were located in the middle of the cuff (under chamber 3), and the measurements were repeated.
Data Analysis
For each measurement series, measurements were obtained during 3 complete inflation/deflation cycles. We believe that during such an inflation sequence, pressure (Pchamber as well as Pinterface) reaches different levels (indicated as A, B, C, and D in Fig. 2), corresponding to the inflation of the different chambers. We calculated an average pressure value for each level from data gathered during at least 15 seconds over 3 inflation/deflation cycles. From the 3 inflation/deflation cycles, we calculated one average cycle (which is displayed in Figs. 2, 3, and 4).
|
|
|
on the calculated mean pressure levels measured during the plateau phase (AD), the preset target pressures (30, 60, 80, and 100 mm Hg), and the pressure measurement locations (14). When the ANOVA test reached statistical significance (
=.05), a Tukey test was used for pair-wise post hoc analysis. Measurements were considered statistically significant at P<.05. Similar analysis was used to assess differences in measurement location on the ellipsoidal model, taking into account the data from the 2 series of measurements. The effect of diameter on Pinterface was determined by performing a 3-way ANOVA at each measurement location, with pressure plateau, preset target pressure, and model diameter as independent factors. | Results |
|---|
|
|
|---|
The same tendency was observed for Pinterface (Fig. 2). The effect of inflation of chambers 2 to 4 was clearly observed in the pressure recordings at location 1. During most of the inflation sequence (inflation chambers 13), the highest pressure was at location 1, with a pressure gradient from the extremity of the limb toward the body. When chamber 4 was inflated, however, pressure at location 3 became the highest, and there was no longer a continuous distal-proximal pressure gradient. Overall, Pinterface had about the same order of magnitude as the corresponding Pchamber. For instance, at location 1 and for the 4 target pressures, maximal Pinterface values were 60, 97, 116, and 140 mm Hg, respectively, and maximal Pchamber values were 54, 98, 121, and 141 mm Hg, respectively.
Analysis of variance revealed that, for both Pchamber and Pinterface, the pressure plateau values, as we expected, were dependent on the target pressure level (P<.001), the measurement location (P<.001), and the plateau identification (AD) (P<.001). This means that, within one inflation sequence and at one measurement site, pressure was different at plateaus A through D, indicating an interaction among chambers. This was the case for all model measurements.
There was little effect of the diameter of the model on Pchamber. Pinterface, in contrast, showed marked effects, as shown in Figure 3, for measurement location 2. The smaller the diameter, the higher the pressure. There was an effect of diameter at measurement locations 1 (P<.001), 2 (P<.001), 3 (P<.05, with the only statistically significant difference between 80 and 100 mm), and 4 (P<.001).
We found an effect of limb geometry and measurement location on the pressure exerted on the cuff-skin interface (Fig. 4). For the ellipsoidal model, there was an effect of measurement location (P<.001). Pressure at measurement locations 3 and 4 (opposite of the long axis; see Fig. 4) were not different, but were higher (P<.05) than pressures at measurement locations 1 and 2 (opposite of the short axis; see Fig. 4). Pressures at measurement locations 1 and 2 were not different from each other.
| Discussion |
|---|
|
|
|---|
The effect of diameter and curvature, we believe, can be explained on the basis of Laplace's law (Fig. 5). Given Pchamber and Pinterface, the local radius (Rl), the stress in the cuff material (
1), and the wall thickness (h), equilibrium of stresses yields Pinterface=Pchamber+
1h/Rl. Thus, theoretically, Pinterface is always higher than Pchamber, except on flat contact surfaces (Rl=
) or when the wall material does not bear any stress (
1=0). Our data are consistent with Laplace's law: Pinterface values were higher for the smaller model diameters and for those measurement locations on the ellipsoidal model with the smallest curvature.
|
1) may be absorbed in the inner cuff wall. The exact value of
1 is difficult to estimate, as it depends on cuff material properties, on friction between the cuff and the model, and on the folding pattern of the cuff. Due to the large size of the infusion bags, measured Pinterface represents an average pressure exerted over a relatively wide area. The disadvantage is that it is difficult to measure strictly underneath individual cuff chambers. We believe that the high pressure measured at location 3 for the 80-mm cylindrical model (Fig. 2) was due to the fact that the infusion bag was not well centered underneath chamber 3, where we might have captured some influence of the stiffer stitching seam. Nonetheless, these high pressures actually occur and may be important, as they can reverse the pressure gradient. Overall, we believe that the disturbing effect of the presence of the infusion bags is limited for the uniform model data. Cuff pressures and measured Pchamber were of the same order of magnitude and followed the same "typical" pattern with a progressive increase in pressure upon inflation of adjacent cuff chambers. The effect is probably more important for the ellipsoidal model, as 4 bags are connected at the same level under the cuff.
Both the Pinterface and Pchamber data showed an interaction between adjacent cuff chambers. This interaction caused the most distal pressure to rise continuously, ultimately reaching values 40% to 80% higher than the preset working pressure for working pressures of 30 to 100 mm Hg. The increase in pressure is explained by simple mechanics (Fig. 5). Upon inflation of the first chamber to pressure P1, the partition between chambers 1 and 2 takes a "rounded" shape, and a steady state, with equilibrium between Pchamber and wall stresses, is obtained. With zero pressure in chamber 2, the transmural pressure difference is P1. Inflating the adjacent chamber to pressure P2, the shape of chamber 1 is more or less maintained, together with the coexisting stresses in the wall. If the shape and stresses are completely maintained, equilibrium requires a constant transmural pressure difference of P1, and pressure in chamber 1 therefore would rise to P1+P2. Measurements show that P2 is not entirely added to P1, suggesting that the shape of chamber 1 is probably not entirely maintained upon inflation of the other chambers.
Cuffs have been designed to generate a more wavelike compression, proceeding from the extremity of the limb toward the heart. This compression mode is much more effective than uniform compression in evacuating fluid.13,14 We argue, however, that it is unacceptable for pressures to rise far above the desired working pressures. There are several possible ways to prevent these excessive pressures. A simple method could be to use different target pressures for each chamber, with lower target pressures for the more proximal chambers. Another solution may consist of better control algorithms in the pressure controller: instead of measuring pressure only during inflation of each chamber, repetitive measurements in each chamber could allow better control of target pressures. Other alternative cuff designs such as those with controllable pressure valves mounted on each cuff chamber might prevent pressures from rising above a desired value.
In clinical practice, compression therapy and lymphatic drainage are used for the prophylaxis of deep venous thrombosis.911,17,18 Target pressures depend on the application, but range from 30 to 100 mm Hg or higher. Especially for the prevention of deep venous thrombosis, working pressures of 100 mm Hg or higher are used.10 The prophylactic effect of pneumatic compression on deep venous thrombosis is attributed to the intermittent application of high flows and shear stresses, therebyat least in theoryalso preventing stasis. Nicolaides et al11 found that pressures higher than 35 mm Hg did not increase venous peak flow in the limb. Generally, for the treatment of lymphatic edema, lower target pressures (around 30 mm Hg) are used.3 Nevertheless, cuff interaction causes the pressures to rise to 50 mm Hg or higher. Such high pressures may be detrimental to the lymphatic circulation.8
Limitations
Our study has some limitations. Only one device was used during the experiments. We believe, however, that the apparatus and its control algorithm are representative for this type of device. The model we used consisted of a simple, rigid structure. We contend, however, that the Pinterface values we recorded are representative of the pressures exerted on the soft tissues of the limb for 2 reasons. First, the perimeter of the inside of the cuff is larger than the perimeter of a limb, and upon inflation, the cuff folds itself around the object inside, creating what we believe is perfect contact, with Pchamber fully transferred to the contact surface. Although this is our contention, we have no data to support this hypothesis. Second, by using fluid-filled bags to measure Pinterface, we believe a deformable soft tissuelike layer is induced, making the cuff-model interface less rigid.
Our measurements do not give information on the pressure distribution inside the limb (which depends on the exact geometry and constitution of the limb) or information on the effect of compression therapy on venous or lymphatic flow. In addition, the reproducibility of the measurements, particularly with respect to Pinterface, depends on the exact positioning of the model inside the cuff. Although we did not do a thorough repeatability study, we believe the observation that there was an interaction among cuff chambers was not attributable to measurement error. The reported data were measured on 4 different models that were taken apart and instrumented between measurement sessions. The typical pressure pattern was present in all measurements (both Pchamber and Pinterface), and there was an interaction. Within one measurement sequence (ie, different inflation/deflation sequences without removing the model), data are highly reproducible. The maximum standard deviation of measured plateau pressure was less than 2 mm Hg. In addition, for the same target pressure level, Pchamber wasas we expectedindependent of cylindrical model diameter, and the anticipated effect of diameter on Pinterface could be demonstrated.
| Conclusion |
|---|
|
|
|---|
| Footnotes |
|---|
Dr Segers is the recipient of a post-doctoral grant of the Fund for Scientific ResearchFlanders (Belgium)
* GymnaUniphy BV, Ekkersrijt 4401, 5292 DL Son, PO Box 558, 5600 AN Eindhoven, the Netherlands. ![]()
Becton-Dickinson Medical Systems, Denderstraat 24, PO Box 13, 9320 Erembodegem-Aalst, Belgium. ![]()
National Instruments Corp, 11500 N Mopac Expressway, Austin, TX 78759. ![]()
Jandel Scientific, SPSS Science, 233 S Wacker Dr, 11th Floor, Chicago, IL 60606-6307. ![]()
|| Entran Devices Inc, 10 Washington Ave, Fairfield, NJ 07004. ![]()
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
H. N Mayrovitz Interface Pressures Produced by Two Different Types of Lymphedema Therapy Devices Physical Therapy, October 1, 2007; 87(10): 1379 - 1388. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |